        • 產品名稱:Vermason 222547校準器

        • 產品型號:用于校準手腕帶和工鞋接地測試儀
        • 產品廠商:其它品牌
        • 產品文檔:
        你添加了1件商品 查看購物車
        簡單介紹:
        Vermason 222547校準器 Calibration Unit for Vermason Wrist Strap and Footwear Tester用于校準手腕帶和工鞋接地測試儀適配儀器:222504, 222506,222507, 222562, 222563, 222564, 222565,Vermason(DESCO子品牌)
        詳情介紹:

        Vermason 222547校準器 Calibration Unit for Vermason Wrist Strap and Footwear Tester
        用于校準手腕帶和工鞋接地測試儀
        適配儀器:222504, 222506,222507, 222562, 222563, 222564, 222565
        Vermason(DESCO子品牌)

        Operation
        The Tester Calibration Unit was specifically designedfor use in calibration of Vermason brand equipment.The following procedures will not cover adjustment of
        the test equipment. For detailed information regardingadjustment of specific Vermason brand testers, pleasecontact Desco Europe Customer Service.

        Frequency of recalibration should be based on thecritical nature of those ESD sensitive items handled andthe risk of failure for the ESD protective equipment andmaterials. In general, we recommend that calibration be
        performed annually.

        Testing the Compact Tester
        WRIST STRAP CIRCUIT
        1. Connect the 10 mm lead from the Calibration Unitto the wrist strap stud on the Compact Tester.
        2. Select 715K FAIL LOW on the Calibration Unit.
        3. Press and hold the Calibration Unit’s test probeon the Compact Tester’s test plate. The testershould indicate a FAIL LOW condition.
        4. Select 785K PASS LOW on the Calibration Unitand repeat the test. The tester should indicate aPASS condition.
        5. Select 33M PASS HIGH on the Calibration Unitand repeat the test. The tester should indicate aPASS condition.
        6. Select 37M FAIL HIGH on the Calibration Unitand repeat the test. The tester should indicate aFAIL HIGH condition.

        FOOTWEAR CIRCUIT
        1. Connect the 10 mm lead from the Calibration Unitto the 10 mm footwear stud on the Compact Tester.
        2. Select 715K FAIL LOW on the Calibration Unit.
        3. Press and hold the Calibration Unit’s test probeon the Compact Tester’s test plate. The testershould indicate a FAIL LOW condition.
        4. Select 785K PASS LOW on the Calibration Unitand repeat the test. The tester should indicate aPASS condition.
        5. Select 33M PASS HIGH on the Calibration Unitand repeat the test. The tester should indicate aPASS condition.
        6. Select 37M FAIL HIGH on the Calibration Unit andrepeat the test. The tester should indicate aFAIL HIGH condition